dLab1

Optical measuring tool for O-rings and technical components, available in 4 sizes.

Laboratory machine  dLab1

Inspections

  • Dimensionals

The inspections that can be carried out are:

Rings

  • Minimum and maximum internal diameter
  • Minimum and maximum external diameter
  • Maximum difference between the maximum and minimum internal diameter
  • Minimum and maximum cord value
  • Average internal diameter
  • Average external diameter
  • Average cord

Irregular Figures/Borders

  • Four programmable rulers for measuring linear distances on the part template, established through automatic calibration.
  • Automatic point measurement system (dedicated software) that makes it possible to obtain orthogonal measurements even for non-parallel contours.
  • Radii of partial curvature, inter-axial distance, internal diameters.
  • Item memorized in order to perform subsequent consecutive measurements without needing to orient the part.
  • Comparison between tolerances (five positive and five negative or zero with respect to the nominal value).
  • Values memorized in a dedicated table with calculation of CP and CPK.

 

Description

dLab 1 is a manual loading/unloading machine that is suitable for contact-less measurement.

Parts are measured using a camera with a resolution of 5 Mpx, equipped with telecentric optics, which makes it possible to acquire distortion-free images.

The data related to this image is transferred to the computer, which makes it available to the operator in numerical form thanks to a graphic interface. This data is then stored in a database and can be accessed at any time. Measurements can be made on both circular and/or irregular shapes like borders.

dLab 1 offers the possibility of taking multiple measurements by positioning several parts within the framed field at the same time; the values obtained will take into account the preset average tolerances.

  • Statistics and Reports
  • Batch Traceability
  • Remote Management
  • Interfacing with Management Software (M.E.S.)

Technical characteristics

Technical Data Sheet

dLab1 S

dLab1 M

dLab1 L

dLab1XL

Framed Field

45 mm

64 mm

130 mm

Ø 225 mm

Camera Resolution

5Mpx

5Mpx

5Mpx

12 Mpx

Max. Measurable Part Diameter

42 mm

60mm

125 mm

200 mm

Min. Measurable Part Diameter

0.5 mm

0.5 mm

1 mm

2 mm

Allowed Part Thickness (Z)

0.5 ÷ 7 mm

0.5 ÷ 15 mm

0.5 - 30 mm

0.5 - 50 mm

Processing Time

1 pz/sec.

1 pz/sec.

1 pz/sec.

1 pz/sec.

Resolution [mm/pixel]

0.022

0.031

0.064

0.075

Measurement Repeatability

± 0.004 mm

± 0.005 mm

± 0.01 mm

± 0.013 mm

Measurement Accuracy

± 0.008 mm

± 0.010 mm

± 0.021 mm

± 0.025 mm

Noise Level

< 15 dB

< 15 dB

< 15 dB

< 15 dB

Single-phase Power Supply Voltage

230V - 50Hz

230V - 50Hz

230V - 50Hz

230V - 50Hz

Operating Temperature

+5°C to +40°C

+5°C to +40°C

+5°C to +40°C

+5°C to +40°C

Relative Humidity

max 80%

max 80%

max 80%

max 80%

Electrical Protection

IP 44

IP 44

IP 44

IP 44

Installed Power

0.5 kw

0.5 kw

0.5 kw

0.5 kw

Accuracy and repeatability are evaluated on the basis of a statistical test, in the range of ±3σ, where σ represents the standard deviation.

Size

Dimensions dLab1

Select your language